LV
Product
Life2000 Ventilation System
by Baxter · last CVE Nov 14, 2024
8
CVEs
0
On KEV
10.0
Max CVSS
0
New 30D
Severity mix
Critical 8High 0Medium 0Low 0
Watch

Add Life2000 Ventilation System to your stack to get only the CVEs that touch what you run.

CVECVSSEPSSPublishedSummary · affected versions
CVE-2024-48967
10.0
1%
Nov 14
Life2000 ventilator and Service PC lack sufficient audit logging capabilities
Affected: 06.08.00.00 and prior
CVE-2024-48966
10.0
1%
Nov 14
Life2000 service tools for test and calibration do not support user authentication
Affected: 06.08.00.00 and prior
CVE-2024-48970
9.3
0%
Nov 14
Life2000 Ventilator microcontroller lacks memory protection
Affected: 06.08.00.00 and prior
CVE-2024-48974
9.3
0%
Nov 14
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Affected: 06.08.00.00 and prior
CVE-2024-48973
9.3
0%
Nov 14
Debug port on Life2000 Ventilator serial interface is enabled by default
Affected: 06.08.00.00 and prior
CVE-2024-48971
9.3
0%
Nov 14
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
Affected: 06.08.00.00 and prior
CVE-2024-9832
9.3
0%
Nov 14
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Affected: 06.08.00.00 and prior
CVE-2024-9834
9.3
0%
Nov 14
Improper data protection on Life2000 ventilator serial interface
Affected: 06.08.00.00 and prior
Showing 8