CVE-2022-4992
Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering
8.6
CVSS
0.2%
EPSS
No
On KEV
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Confidentiality
Low
Integrity
Low
Availability
High
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.2% · higher than 5% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
IA
Infinity Acute Care System · Dräger
before VG4.2; before VG4.1.1; before VG4.0.3 · source: vendor
SI
Standalone Infinity M540 patient monitor · Dräger
before VG4.2; before VG4.1.1 · source: vendor