High severity · published Jun 2, 2026
CVE-2019-25719
Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
8.6
CVSS
0.1%
EPSS
No
On KEV
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Confidentiality
Low
Integrity
Low
Availability
High
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.1% · higher than 3% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
IA
Infinity Acute Care System · Dräger
through VG4.1.1; through VG4.0.3 · source: vendor
SI