Description
In Patient Information Center iX (PICiX) Versions C.02, C.03,
PerformanceBridge Focal Point Version A.01, the product receives input
that is expected to be well-formed (i.e., to comply with a certain
syntax) but it does not validate or incorrectly validates that the input
complies with the syntax, causing the certificate enrollment service to
crash. It does not impact monitoring but prevents new devices from
enrolling.
CVSS vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L
Attack vector
Adjacent
Complexity
Low
Privileges
None
User interaction
None
Confidentiality
None
Integrity
None
Availability
Low
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.3% · higher than 20% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
PI
Patient Information Center IX · Philips
b.02; c.02; c.03 · source: NVD
PF
Performancebridge Focal Point · Philips
a.01 · source: NVD
References
NVD
https://nvd.nist.gov/vuln/detail/CVE-2020-16220
CVE.org
https://www.cve.org/CVERecord?id=CVE-2020-16220
Third Party Advisory
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01
www.philips.com
https://www.philips.com/productsecurity
More from Patient Information Center IX
8.8
CVE-2020-16222
In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software…
6.8
CVE-2020-16212
In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate…
6.5
CVE-2021-43550
The use of a broken or risky cryptographic algorithm is an unnecessary risk that may result in the exposure of sensitive information, which affects the communications between…
6.5
CVE-2021-43548
Patient Information Center iX (PIC iX) Versions C.02 and C.03 receives input or data, but does not validate or incorrectly validates that the input has the properties required to…
6.5
CVE-2020-16224
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that…