High severity · published Jun 29, 2020
CVE-2020-12048
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and…
7.5
CVSS
0.5%
EPSS
No
On KEV
Description
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Confidentiality
High
Integrity
None
Availability
None
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.5% · higher than 39% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
PX
Phoenix X36 Firmware · Baxter
3.36; 3.40 · source: NVD