High severity · published Jun 1, 2026
CVE-2019-25718
Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass
7.8
CVSS
0.1%
EPSS
No
On KEV
Description
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
Attack vector
Local
Complexity
Low
Privileges
Low
User interaction
None
Confidentiality
High
Integrity
High
Availability
High
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.1% · higher than 2% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
IE
Infinity Explorer C700 · Dräger
Infinity Explorer C700 · source: vendor
IE
Infinity Explorer C700 Firmware · Draeger
All versions, or not specified · source: NVD
References
NVD
https://nvd.nist.gov/vuln/detail/CVE-2019-25718
CVE.org
https://www.cve.org/CVERecord?id=CVE-2019-25718
Vendor Advisory
https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf
Third Party Advisory
https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypass