Medium severity · published Aug 13, 2018
CVE-2018-14781
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
5.3
CVSS
0.7%
EPSS
No
On KEV
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
CVSS vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Attack vector
Adjacent
Complexity
High
Privileges
None
User interaction
None
Confidentiality
None
Integrity
High
Availability
None
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.7% · higher than 53% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
M5
MMT – 511 pump Paradigm · Medtronic
All versions · source: vendor
M5
MMT- 508 - MiniMed pump · Medtronic
All versions · source: vendor
5M
508 Minimed Insulin Pump Firmware · Medtronicdiabetes
All versions, or not specified · source: NVD
5P
522 Paradigm Real-time Firmware · Medtronicdiabetes
All versions, or not specified · source: NVD