Medium severity · published Feb 13, 2017
CVE-2016-8375
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit
4.9
CVSS
1%
EPSS
No
On KEV
Description
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device's flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.
CVSS vector
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N
Attack vector
Physical
Complexity
High
Privileges
None
User interaction
None
Confidentiality
High
Integrity
None
Availability
None
Scope
Changed
Exploitation
Is it being used?
Exploit prediction (EPSS)1.0% · higher than 63% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
A8
Alaris 8015 PC Unit · BD
through 9.5; 9.7 · source: NVD
Affected product